Verification Procedure; Hardware Set-Up; Software Set-Up - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

236
11.7.8

Verification procedure

To verify the correct operation of the Master/Slave configuration, proceed as
follows:

Hardware set-up

1
Set-up two GEN series mainframes with each a Master/Slave card installed
and with each at least one recorder card installed.
2
Connect a TTL level, 1 Hertz signal to the top input of the first recorder card
of the master mainframe and to the top input of the first recorder card of
the slave mainframe.
3
Switch on both GEN series mainframes and wait until they have completed
the booting process.
4
With the fiber optic cable, connect to any connector of the Master/Slave
card of the master mainframe to the top connector labelled M/S IN of the
Master/Slave card of the slave mainframes.
5
Check if both LEDs on both Master/Slave cards are illuminated green.

Software set-up

1
If it is not already active, start Perception.
2
In the start dialog, select New blank experiment.
3
Make sure that you are connected to the required mainframes. Use the
Hardware Navigator to do this.
4
In the Settings sheet, go to the General group in the task pane and select
Mainframe. A list of available mainframes is displayed in the settings area.
5
Set the master mainframe operating mode to Master in the Master/
Slave mode column.
6
Set the slave mainframe operating mode to Slave in the Master/
Slave mode column.
7
The slave mainframe will now be synchronized to the master mainframe.
The status palette will now show a box with the synchronization status of
the Master/Slave system. This box is labelled MASTER SLAVE.