Using the 1 Gbit Option Connections

IM2 - Master/Slave Synchronization

11.2.2
The IM2 module has built-in synchronization capability for GEN2i systems only.
The following table shows the various supported configurations with the IM1,
IM2 and Master/Slave (M/S) cards installed.
IM2 - Master/Slave Synchronization Support
GEN16t, GEN7t,
GEN5i
(16t, 7t, 5i)
GEN2i
Legend:
Not supported.
Standard Synchronization: Synchronizes the first sample in the
recording for each mainframe, prevents frequency drift of the sample
rates within each mainframe, synchronously exchanges every channel
trigger connected to the Master/Slave trigger bus to/from each
connected mainframe and automatic cable length delay compensation.
LC Connection Using the SFP
+ Option
LC optical connections that
need an SFP device to enable
their use with LC connected
optical cable.
Master Slave Synchronization
= Link up/
active(green)
On; Link up, correct
Synchronized Recording
(M/S) link, fiber cable
connected
Off; Link down, No Fiber
cable connected
Flashing; Fiber cable connected, on-board
Synchronized Recording (M/S) disabled (Disabled
by user or by present Master/Slave card).
IM2
IM1 + Master/
Slave card
GEN5i
= Receiver enable
(green)
On; Active
Off; Inactive
IM2 + Master/
Slave card
147