General Specifications
Analog Input Section
Component
(1)
IEPE amplifier support
Ranges
Excitation current
Excitation accuracy
Coupling time constant
Current Shunt Support
Ranges
(2)
Accuracy
Measurement Shunt
Maximum Current
Overload Protection
IEPE refers to internally amplified sensors - low impedance, piezoelectric force, acceleration and pressure type sensors with built-in
(1)
integrated circuits.
(2)
Errors are listed for amplifier with filter (IIR or FIR)
Analog to Digital Conversion
Component
Model
Sample rate
ADC resolution
Time base accuracy
Bessel or Butterworth filter specifics
Analog anti-aliasing
Time Domain
Frequency Domain
IIR or FIR filter specifics
Digital
Frequency domain
Time domain
On-board Memory
Component
Model
Per card (Mega Samples)
Per channel
I2679-4.0 en
Unit Description
7 ranges from in 1, 2, 5 steps
Software selectable in 1 mA steps
Nominal
5 ranges in 1, 2, 5 steps
Resettable fuse, 0.1 Ω ± 20 %
Universal 200 ISO CARD
200 kS/s to 0.1 S/s
Time- or Frequency domain optimized
20 kHz
7-pole Butterworth, extended frequency response
20 kHz
12-pole FIR at sample rate divided by:
4, 10, 20, 40
6-pole Bessel style IIR, sample rate divided by:
10, 20, 40, 100
Universal 200 ISO CARD
64 MS shared by enabled channels
16 MS per channel
Value
16 bit (0.0015 %)
50 ppm
Bypass, Time, Frequency-domain optimized
7-pole Bessel, optimal step response
IIR or FIR
12-pole FIR at sample-rate divided by:
6-pole Bessel style IIR, sample rate divided by:
Value
256 MS shared by enabled channels
GEN5i
Value
± 0.2 V to ± 20 V
1 to 15 mA
24 V
1 s
± 50 mA to ± 1 A
≤ 0.2 % of FS } ± 300 μA
0.2 Ω ± 1 %
1 A
1.6 A
Universal 1M ISO CARD
1 MS/s to 0.1 S/s
< ± 0.2 V: 185 kHz (-3 dB)
≥ ± 0.2 V: 220 kHz (-3 dB)
< ± 0.2 V: 300 kHz (-3 dB)
≥ ± 0.2 V: 350 kHz (-3 dB)
4, 10, 20, 40
10, 20, 40, 100
Universal 1M ISOCARD
64 MS
477