How to use the 5B based amplifier with Genesis and Perception
As an example of a working 5B Integration card we will consider a card with
two 5B modules mounted, one being an f to V converter and one being a TC
module. These two have fixed ranges like 0-10.000 Hz and -200 °C to
+600 °C, each being 0-5 Volt at the output.
Connect the output (preferable BNCs mounted on the 5B Integration card front
panel) to two selected Genesis input channels now acting as "Frequency" and
"TC" channel. Set sensitivity in these channels to 0-5 Volt (or whatever the 5B
modules used are delivering as full scale value).
Scale the inputs to 0-10,000 Hz and -200 °C to +600 °C using the TU (technical
units) input fields for those channels in Perception.
In order to do this either calculate the TU Multiplier value, or just do a two-point
calibration and enter the upper and lower values there.
Save this setup as hardware setting and start any future measurements from
there.
Figure 12.34: 5B Integration card setup
CH 1A and CH 2A are scaled to measure frequency and temperature using 5B
modules in front of standard input channels.