246
12.2.2

GN611, Isolated 1 kV 200kS/s input card

6 analog channels
l
Isolated, balanced differential
l
inputs
± 20 mV to ± 1000 V input range
l
600 V RMS CAT II isolation
l
User selectable digital Bessel,
l
Butterworth and Elliptic filters
200 kS/s sample rate
l
18 bit resolution
l
200 MB memory
l
Two 4 mm banana plugs for
l
each channel
Real time cycle based
l
calculation; Triggering on
calculated result
Digital Event/Timer/Counter
l
support
The isolated balanced differential
inputs allow direct connection to any
voltage up to 600 V RMS on both
inputs of each channel.
The signal conditioner provides
voltage inputs from ± 20 mV to
± 1000 V combining low and high
voltage input ranges all in one card.
The model uses two 4 mm safety
banana plugs for each channel. A standard safety rated BNC-to-banana
adapter can be used to easily connect coax cables using BNCs.
This card meets the international IEC61010-1 safety standard. The card offers
600 V RMS CAT II and 300 V RMS CAT III isolation to allow safe measurements
within the most demanding electrical environments.
Every channel is equipped with an independent full range input amplifier, 7-pole
Bessel or Butterworth analog anti-aliasing filter, user selectable digital Bessel,
Butterworth and Elliptic IIR filters and an 18-bit Analog-to-Digital converter.
For true real time analysis the card offers real time cycle based calculations.
Automatic zero crossing detection allows for asynchronous true RMS, mean
and other calculations that can be used to trigger the recording.
If supported by the selected mainframe, the GEN DAQ series input card offers
16 digital input events, two digital output events and two timer/counter
channels.
For specification and ordering information, please refer to "B3716-2.0 en (GEN
series GN611)" on page 381.
GEN5i
I2679-4.0 en

Advertisement

Table of Contents
loading