HBM SomatXR CX23-R User Manual

HBM SomatXR CX23-R User Manual

Somatxr series ultra-rugged daq
Hide thumbs Also See for SomatXR CX23-R:
Table of Contents

Advertisement

User Manual
English
CX23-R and EX23-R
SomatXR Ultra-Rugged DAQ

Advertisement

Table of Contents
loading

Summary of Contents for HBM SomatXR CX23-R

  • Page 1 User Manual English CX23-R and EX23-R SomatXR Ultra-Rugged DAQ...
  • Page 2 19 Bartlett Street Marlborough, MA 01752 Tel. +1 800-578-4260 info@usa.hbm.com Mat.: DVS: a4275-8.0 en HBM: public 03.2018 © HBM, Inc. Subject to modifications. All product descriptions are for general information only. They are not to be understood as a guarantee of quality or durability.
  • Page 3: Table Of Contents

    4.1   Setting up a test 4.2   Configuring input channels 4.2.1   Configuring TEDS sensors 4.2.2   Vector CAN databases 4.2.3   eDAQ CAN databases 4.2.4   HBM sensor databases 4.3   Input channels 4.3.1   Input channel generic parameters 4.3.2   Input channel scaling parameters 4.3.3   Analog channels 4.3.4   CAN channels...
  • Page 4 4.6.1   DataMode™ generic parameters 4.6.2   Burst history 4.6.3   Burst message logger 4.6.4   Event slice 4.6.5   Message logger 4.6.6   Peak valley 4.6.7   Peak valley matrix 4.6.8   Rainflow 4.6.9   Time at level (One Dimensional) 4.6.10   Time history a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 5 5.4.6   Current-fed piezoelectric transducers (IEPE, ICP®) 5.4.7   Current sources 5.4.8   Resistance-based measurements 5.4.9   Thermocouples 5.4.10   Digital inputs 5.4.11   Digital outputs 5.4.12   Digital timer inputs 5.4.13   Passive inductive encoder 5.4.14   Pulse width, pulse duration, period duration (PWM) 5.4.15   CAN devices CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 6 6.9   Test and data control 6.9.1   Test control 6.9.2   SIE data 7   Operational notes 7.1   System timing 7.2   Data synchronization 7.3   Error handling 7.4   Long term unattended testing considerations 7.5   Troubleshooting 7.6   Reset to factory defaults rescue a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 7 7.7   Shunt scaling based on a known shunt scaling 7.8   SIE file metadata 7.9   SIE file size issues CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 8: Safety Instructions

    The supply connection, as well as the signal and sense leads, must be installed in such a way that electromagnetic interference does not adversely affect device functionality. (HBM recommendation: "Greenline shielding design," downloadable from the Internet at www.hbm.com.) Automation equipment and devices must be covered over in such a way that adequate protection or locking against unintentional actuation is provided (such as access checks, password protection, etc.).
  • Page 9   The device is not used in accordance with the operating manual.   The device is used outside the field of application described in this chapter.   The operator makes unauthorized changes to the device. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 10 In particular, any repair or soldering work on motherboards or replacement of components is prohibited. When exchanging complete modules, use only original parts from HBM. The module is delivered from the factory with a fixed hardware and software configuration. Changes can only be made within the possibilities documented in the manuals.
  • Page 11 CAUTION Meaning: Potentially dangerous situation Warns of a potentially dangerous situation in which failure to comply with safety requirements could result in bodily injury or damage to property. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 12 Safety instructions NOTE Meaning: Important information Points out important information about the product or its handling. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 13: Getting Started

    MX460B-R Frequency / Counter Module: Up to four (4) digital inputs to measure digital pulses up to 1 MHz (speed, torque, angle position, displacement, PWM in general). CX23-R/eDAQXR support is limited in the following aspects. Math CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 14 Interactive zeroing of the “crank shaft” sensors is not supported.   MX471B-R CAN Module: Up to four (4) independent CAN bus nodes electrically isolated from each other and from the power supply. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 15: Module Transducer Overview

        ●          ●          ●        (IEPE, ICP®) ● ● ● Current,         20 mA Resistance ● ●           or resistance-based measurements CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 16 Use quarter bridge adapter 1-SCM-R-SG1000-2, 1-SCM-R-SG120-2 or 1-SCM-R-SG350-2. † Use ODU 14-pin to BNC adapter 1-KAB430-0.3. ‡ Use thermocouple adapter 1-SCM-R-TCK-2 for K-type, other types need a dedicated adapter. § Including support for CCP/XCP-on-CAN (not in combination with CX23-R/EXRCPU). a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 17: Cx23-R Data Processor

    1.2.1   Setting up a basic system To set up a CX23-R module with basic connectivity, perform the following steps: CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 18 15 seconds for the unit to boot up.  6. Open a web browser and navigate to the default IP address 192.168.100.101. If presented with a certificate warning, choose to proceed. The web interface is displayed. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 19: Configuring And Running A Basic Test

    Use the EX23-R Ethernet Switch to network SomatXR modules with other devices such as cameras. Use the power over Ethernet (PoE) ports to provide power to compatible network devices. The EX23-R is designed to be used straight out of the box. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 20  2. Connect the desired devices to the Ethernet ports using 1-KAB2100, 1-KAB2106 or 1-KAB2107 Ethernet cables, depending on the device. CAUTION Drawing PoE power lowers the maximum operating temperature. Please refer to the data sheet for the exact de-rating specifications. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 21: Setting Up The Hardware

    MX module to the X101 or X102 connector on the second MX module.  3. Connect the CX23-R ETHERNET1 and ETHERNET2 connectors to the MX module X100 connectors using KAB2100 cables. This establishes communication and synchronization between the CX23-R and MX modules. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 22 EX23-R module to any open EX23-R Ethernet port.  3. Use 1-KAB2100 cables to connect the MX module X100 connectors to the EX23-R switch. MX module firmware update process Do not reboot the CX23-R when MX firmware updates are in progress. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 23: Setting Up Somatxr Wireless

    2   Setting up the hardware 2.2   Setting up SomatXR wireless HBM offers a wireless solution for the SomatXR system using the Sierra Wireless GX Series, MOXA AWK-4121 or ACKSYS WLg-xROAD devices. SIERRA WIRELESS GX Series gateway The Sierra Wireless GX Series cellular gateway is capable of 4G LTE network connection.
  • Page 24  9. Exit the web interface.  10. Disconnect the computer from the EX23-R.  11. Connect the EX23-R to the access point and the CX23-R as shown in Connecting the access point for Power over Ethernet. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 25 Connecting the access point to a CX23-R The ACKSYS WLg-xROAD/N and ACKSYS WLg-xROAD/NP access points can be connected directly to the CX23-R, but the access points require connection to a 9–50 V power source. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 26  9. Exit the web interface.  10. Disconnect the computer from the EX23-R.  11. Connect the EX23-R to the access point and the CX23-R as shown in Connecting the access point for Power over Ethernet. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 27: Setting Up An Axis Network Camera

     6. Use a 1-KAB2107-X cable to connect the CX23-R ETHERNET 1 port to EX23- R 2 port. 2.3   Setting up an Axis network camera The SomatXR/eDAQXR system offers camera input channels through the use of Axis network cameras with the EX23-R PoE Switch. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 28   Axis Q1775 Network Camera   Axis P1214-E HD Network Camera NOTE HBM only supports the devices listed above and assumes no risk if any other devices are used. Resetting the camera to factory defaults For best results or if issues exist with detecting the camera, reset the camera to factory defaults.
  • Page 29: Setting Up A Remote Power Switch

    SomatXR system power source. The included diagrams are not intended to be complete, detailed instructions. CAUTION Connection to the positive power terminal without proper grounding may result in a blown fuse and/or other damage to the SomatXR system. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 30 This method, while feasible, is not recommended and cannot be guaranteed as safe. Results may include unwanted multiple runs of data, improper reboots, lost data due to multiple power cycles and improper charging of the internal battery pack. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 31 CAUTION Failure to follow these suggestions may result in blown fuses and/or permanent damage to SomatXR modules. Improper powering of the unit requiring repairs by HBM technicians may be deemed as non-warranty usage, resulting in service charges. CX23-R and EX23-R...
  • Page 32 2   Setting up the hardware a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 33: Configuring The System

    If on a network with a DHCP server, use the HBM Device Manager (on the included USB drive) or contact your IT administrator to find out the IP address. If presented with a certificate warning, choose to proceed.
  • Page 34: Managing Users And Privileges

    The unit comes factory set with one user account and one profile with complete administrator privileges. This default user comes with the credentials:   User name: admin   Password: password It is strongly recommended to change this password. The default user and profile cannot be deleted. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 35: Configuring Mx Modules For Use With The Cx23-R

    Changing the sample rate domain The MX module sample rate domain can be either Decimal or Classic and must match that of the test setup, which is based on the System preferences when the test setup is created. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 36: Updating Main Processor Firmware

    To find the exact sample rate, multiply the nominal sample rate by 8388608/8386425. 3.4   Updating main processor firmware HBM regularly releases new firmware introducing new features and providing device improvements. CAUTION Allow the firmware upgrade to run completely. Before upgrading, make sure that no other users are logged in and no other processes are running.
  • Page 37: Setting Up And Running A Test

    CAN databases. For applicable transducer types, sensor parameters can be easily defined using TEDS or imported HBM sensor database files. For channels that need to be scaled, set the scaling parameters and perform experimental scaling measurements.
  • Page 38: Vector Can Databases

    Preparing an HBM sensor database To import an HBM sensor database file, a .sdbx text file must be created from the standard .sdb file using the MX Assistant. To create the .sdbx file:  1. Open the MX Assistant and navigate to the Sensor database tab.
  • Page 39: Input Channels

    4   Setting up and running a test To apply sensor configurations from imported HBM sensor databases, first expand the Sensors panel. The Sensors panel contains the imported sensor databases and their sensor definitions. For very large databases, it may take several seconds to display the database contents for the first time.
  • Page 40: Input Channel Scaling Parameters

    Electrical 2 Physical 2 Experimental two point Experimental two point scaling Internal shunt resistor Internal shunt resistor scaling Strain gage factor Strain gage Bridge factor Scaling parameters   Scaling mode: Set the channel scaling mode. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 41 > 900 or <-450. If the user changes the Range min to -600, the Min alarm threshold changes to -540. However, the Max alarm threshold does not change (i.e., it is still 900). CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 42 In brief, it provides feedback to the user on the accuracy of the shunt scaling experiment. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 43 The define know shunt option calculates the equivalent Physical span from a known shunt. (Page 44) and This shunt tool is only applicable when the shunt scaling CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 44: Analog Channels

    BF = Bridge factor (default is 1.0) ES = Equivalent strain 4.3.3   Analog channels Analog channels include Strain gage transducers, Inductive transducers, Piezoresistive transducers, Potentiometric transducers, Voltage sources, Current- fed piezoelectric transducers (IEPE, ICP®), Current sources Resistance-based measurements. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 45 AC.   MX411B-R Highly Dynamic Universal Module channels, FIR filters are not compatible with the AC excitation mode.   Excitation range: Set the regulated excitation voltage range. The available values depend on the module. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 46: Can Channels

    R/eDAQXR CAN interface. They are not currently supported by the MX471B-R or MX840B-R CAN interfaces. The MX471B-R and MX840B-R support up to 128 database channels for each CAN connector. There is no limit on the number of database channels for unit CAN a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 47 SAE J1939 messages with a "null" source address value (0xFE) cause the unit to generate a mask which ignores the source address portion of the CAN message ID. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 48: Gps Channels

    All modules provide GPS PPS time sync pulses that allow the SomatXR/eDAQXR system to align the GPS channel data to be synchronized in phase with other SomatXR/eDAQXR channels (i.e., analog and digital input channels). a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 49: System Channels

    Temperature channels are sourced from a MX1609KB-R Thermocouple Amplifier. NOTE MX840B-R thermocouple channels using a thermocouple adapter (e.g., 1-SCM- R-TCK-2) are included in Analog channels.   Sensor input mode: Value is Thermocouple.   Thermocouple type: Value is K. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 50: Message Channels

      Rotation: Select a rotation angle as desired to show in the channel display from the device. Refer to the device manufacturer's documentation to verify support for rotation; some devices are limited to 0 and 180 degree rotation. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 51: Computed Channels

      Collect: Select the Collect checkbox to write the channel data to the SIE data file during a test run.   Enable notifications: Select the checkbox so notifications (through email or the HBM Push application) may be sent for the Over range detector Computed channel. See Notifications for configuring email and HBM Push notifications to be sent from the system.
  • Page 52: Anomaly Detect

    If the difference between current window mean and reference mean exceeds the specified gate, the channel adds two (2) to the output.   Limit detect a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 53: Bitmap Trigger

    Measurand and Units are inherited from the selected velocity input channel. Description and Range values are initialized on selection of the velocity input channel.   Velocity channel: Set the velocity input channel. In the edit dialog, the channel sample rate is displayed in Hz. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 54: Down Sampler

      Waveform: Set the shape of the output waveform to sine, triangle (ramp) or square.   Duty cycle: For ramp and square type waveforms, specify the duty cycle in percent (1 to 99). Note that for a ramp waveform, a duty cycle of 99 results in a a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 55: Profile Generator

    The decay value is distributed over the cycle period for a uniform decay. 4.4.7   Profile generator This computed channel is provided for HBM internal use only at this time. 4.4.8   Integrator The Integrator computed channel generates an output that is the integral of the input channel.
  • Page 56: Interactive Trigger

    Digital input channels. The system scans the logical input channel for falling edge and rising edge transitions. The pulse period is updated based on the time between rising-to-rising edges and falling-to-falling edges. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 57: Run Stopper

     2. Add a Signal calculator channel with the expression elapsed_time >= 600, where 600 seconds is the desired test run duration.  3. Add a Run stopper channel with the Signal calculator channel as the input channel. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 58: Signal Calculator

    FALSE TRUE if a is FALSE; else FALSE && a && b TRUE if a and b are TRUE; else FALSE a || b TRUE if either a or b are TRUE; else FALSE a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 59 ) The hyperbolic sine of a cosh( a ) The hyperbolic cosine of a cosh tanh tanh( a ) The hyperbolic tangent of a pi() Function used to return Pi {3.141592653589793} CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 60: Smoothing Filter

    For example, for a filter length of five, the filter averages the current sample, the two samples before and the two samples after. Note that the channel backfills the initial output samples with the first fully filtered a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 61: State Mapper

    Min, Max and Value entries separated by a space or tab. Importing state mapper conditions deletes any manually entered values. The conditions are shown in the grid or text editor after import.   Click Clear to clear all state mapper values. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 62: Statistics

    A Time base shifter computed channel defined with a Lag of 1 sample can be used to change this if desired. Statistical math equations Following are the equations used to compute the statistical values where n is the a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 63: Time Base Shifter

      Trigger start mode: Set the start condition for the trigger.   When true: Start when the input channel is TRUE.   On false - true edge: Start when the input channel transitions from FALSE to TRUE. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 64: Track

      Channel: Set the trigger channel used to latch input channel data. By default, the channel latches the input data value when the trigger channel transitions from FALSE to TRUE and holds this data value until the trigger channel transitions a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 65: Triggered Zero Suppression

      Factor: Enter an integer greater than zero as the up sample factor. This has the effect of increasing the sample rate by the product of this integer factor. The following example has a integer factor of three. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 66: Output Channels

    TRUE (i.e., channel value is 1) when a “Set sink” option is used. Use the “Clear sink” to invert this control logic. Select one of the “latched” options to have the output stay in the same state after a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 67: Datamodes

     On false – true edge: Storage of a burst record will be triggered when the trigger channel transitions from FALSE to TRUE.   Trigger channel: Select the desired logical input channel for burst triggering. The trigger channel must have the same sample rate as the input channels. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 68: Burst Message Logger

      Trigger channel: Select the desired logical input channel for burst triggering. Trigger time   Pre (secs): Specify the time period in seconds of data storage before the trigger.   Post (secs): Specify the time period in seconds of data storage after the trigger. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 69: Event Slice

    The Event Slice DataMode provides a channel select tab for both Master channels and Slave channels. 4.6.5   Message logger The Message logger DataMode™ stores one or more message or video channels in the output data file. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 70: Peak Valley

    A floating point value followed by the ‘%’ character: The value must be in the range of 0.0 to 100.0. For each channel the hysteresis gate is computed using the equation: (pv_hysteresis * (range_max – range_min) / 100.0). a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 71: Peak Valley Matrix

      Trigger: Data storage starts when the trigger channel becomes TRUE and continues until the test run stops.   Gate: Data storage occurs when and only when the trigger channel is TRUE. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 72: Rainflow

    The resulting peak valley stream runs through the rainflow cycle counting algorithm to yield the set of closed a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 73   Histogram bounds: Enter the desired Low bound and High bound for each selected input channel. If the range min and max parameters are defined for an input channel, those values are used as the default Low bound and High bound values for the histogram. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 74: Time At Level (One Dimensional)

      Trigger: Data storage starts when the trigger channel becomes TRUE and continues until the test run stops.   Gate: Data storage occurs when and only when the trigger channel is TRUE. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 75: Time History

    DataMode and click the appropriate arrow. Alternatively, drag and drop the desired channels to the appropriate list. 4.7   Running a test Running a test and displaying live data is done from the Test control page of the Test and data control section. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 76: Personal Computer Support Applications

    Installing XR Download Manager  1. Save the XR Download Manager executable (*.exe) file to the computer from the USB flash drive delivered with the main processor or from the www.hbm.com website.  2. Double-click the file icon to install XR Download Manager.
  • Page 77: Somatxr Hardware

    ±60V. They are designed to drive LEDs, relays and other indicators and control at maximum update rates of 20 Hz nominally. The DIO power supply provides power up to 3 W at user-selectable voltages of 5 V or 12 V (within ±5% accuracy). CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 78: Cx23-R Status Leds

    Test run is in progress. ERROR (red) No error. Unit error. View the System logs for information on the error and to turn off the LED. 1 Hz Less than 2 MB of storage remain. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 79 LED sequence for reboot from GUI   Blue Green Yellow 1. Reboot issued. 4 Hz P† 2. System loaded. S‡ S‡ S‡ S‡ 3. System ready. P† P† † Previous state or state dependent ‡ Sequence with other LEDs CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 80: Cx23-R Input Pwr And Aux Pwr Pin Assignments

    Black/White Input 1 White Ground Black/Green Input 2 Green pin side of cable Ground Black/Blue Input 3 Blue Ground Black/Yellow Output 2 Yellow Ground cable detect Black/Brown jumper to pin 1 Output 1 Brown a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 81: Cx23-R Gps Pin Assignments

      Over temperature   Over voltage or over current input power   Internal power supply not working   Ethernet LED Description No link Green Gigabit link is up Orange 10 or 100 Mb link is up CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 82: Ex23-R Power Pin Assignments

    MXB module to any other MXB module using FireWire without both having an Ethernet connection to the data processor can result in acquired data not being synchronized to the data processor on one or more MXB modules. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 83 In addition, a comprehensive set of Butterworth IIR and Bessel IIR digital filters are available for use with the Decimal sample rates. The data processor also supports the traditional (Classic) HBM sample rates and IIR digital filters.
  • Page 84: Mx840B-R Universal Module

    Strain gage, half bridge ● five-wire configuration Strain gage, quarter bridge     ●    three- or four-wire configuration 3-wire only ● Inductive full bridge ● Inductive half bridge LVDT ● (linear variable differential transformer) a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 85 Use quarter bridge adapter 1-SCM-R-SG1000-2, 1-SCM-R-SG120-2 or 1-SCM-R-SG350-2. † Use ODU 14-pin to BNC adapter 1-KAB430-0.3. ‡ Use thermocouple adapter 1-SCM-R-TCK-2 for K-type, other types need a dedicated adapter. Status LEDs The following table gives the descriptions for all LED states. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 86 CAN bus error; CAN error (incorrect parameterization, interface in Bus OFF state; connection error, invalid TEDS CAN data cannot be data) received or processed MX840B-R pin assignments Connect sensors via the 14-pin ODU MINI-SNAP connectors. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 87: Mx1615B-R Bridge Module (Strain Gage Amplifier)

    Up to 16 individually configurable inputs can be connected to the MX1615B-R, including strain gage-based transducers, standardized voltage and resistance or resistance-based measurements (for example, a PT100 resistance thermometer).   Transducer MX1615B-R Wiring diagram Strain gage, full bridge ● six-wire configuration CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 88 System is error free System error Orange System is not ready; boot procedure is running Orange flashing System is not ready; download is active MX1615B-R pin assignments Connect sensors via the 14-pin ODU MINI-SNAP connectors. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 89: Mx1601B-R Standard Module (Universal Amplifier)

    (20 mA) or current-fed piezoelectric transducers (IEPE, ICP®) can be connected to the MX1601B-R module.   Transducer MX1601B-R Wiring diagram Voltage, ● 60 V, 10 V, 100 mV Current-fed piezoelectric transducer †     ●    (IEPE, ICP®) CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 90 System is error free System error Orange System is not ready; boot procedure is running Orange flashing System is not ready; download is active MX1601B-R pin assignments Connect sensors via the 14-pin ODU MINI-SNAP connectors. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 91: Mx1609Kb-R Thermocouple Amplifier

    The current limitation switches the transducer excitation off if current consumption is higher. 5.3.4   MX1609KB-R Thermocouple Amplifier Up to 16 type K thermocouples (NiCrNi) can be connected to the MX1609KB-R module for measuring temperatures.   Transducer MX1609KB-R Wiring diagram ● Thermocouple K-type only Status LEDs CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 92: Mx411B-R Highly Dynamic Universal Module

    (20 mA) or current-fed piezoelectric (IEPE, ICP®) and piezoresistive transducers.   Transducer MX411B-R Wiring diagram Strain gage, full bridge ● six-wire configuration Strain gage, half bridge ● five-wire configuration Strain gage, quarter bridge     ●    three- or four-wire configuration 3-wire only a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 93 The following table gives the descriptions for all LED states. System LED Description Green System is error free System error Orange System is not ready; boot procedure is running Orange flashing System is not ready; download is active   CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 94 Yellow/Black Measurement signal (+) White Measurement signal (-) Active sensor supply (-) Brown Active sensor supply (+) Yellow Sense lead (-) Grey Sense lead (+) Green TEDS (-) Grey/Black TEDS (+) Green/Black Shield Shield a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 95: Mx460B-R Frequency / Counter Module

    Supplying 12-volt transducer power to a 5-volt encoder may damage it. The measurement channels are electrically isolated from the power supply and the interfaces. When TEDS or T-ID is used, the measurement channel is automatically parameterized after connection. Status LED CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 96 Pin 4 and Pin 5 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection! Connect sensors via the 14-pin ODU MINI-SNAP connectors. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 97: Mx471B-R Can Module

    CAN bus     ●    database or raw message database only § Including support for CCP/XCP-on-CAN (not in combination with CX23-R/EXRCPU). Each CAN bus node is electrically isolated from each other and from the power supply. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 98 Intermittent bus errors (warning); no activity on CAN Bus error; CAN interface in Bus OFF status   Channel LEDs Description Green Channel is ready Yellow Boot procedure is running Yellow flashing Download is active Channel has errors MX471B-R pin assignments a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 99: Inputs And Wiring Diagrams

    Strain gage, half bridge ● ● ● five-wire configuration Strain gage, quarter bridge     ●        ●    ● three- or four-wire configuration 3-wire only 3-wire only Use quarter bridge adapter 1-SCM-R-SG1000-2, 1-SCM-R-SG120-2 or 1-SCM-R-SG350-2. Full-bridge strain gage CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 100 Half-bridge strain gage NOTE For half three-wire bridge configurations, a connection between the sense lead and excitation lines must be made in the connector. Quarter-bridge strain gages a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 101 5   SomatXR hardware NOTE The quarter-bridge, three-wire configuration with sense line can be used with the CX23-R, catman®AP or MX Assistant, which support up scale shunting. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 102: Inductive Transducers

    CX23-R, which supports down scale shunting . It is not supported by catman®AP or MX Assistant. 5.4.2   Inductive transducers   Transducer MX840B-R MX411B-R ● ● Inductive full bridge ● ● Inductive half bridge LVDT ●   (linear variable differential transformer) a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 103 5   SomatXR hardware Inductive full-bridge CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 104 5   SomatXR hardware Inductive half-bridge a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 105: Piezoresistive Transducers

    5   SomatXR hardware Linear variable differential transformer (LVDT) 5.4.3   Piezoresistive transducers   Transducer MX840B-R MX411B-R ● ● Piezoresistive transducer CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 106: Potentiometric Transducers

    5   SomatXR hardware 5.4.4   Potentiometric transducers   Transducer MX840B-R MX1615B-R ● ● Potentiometric transducer a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 107 5   SomatXR hardware CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 108: Voltage Sources

    60 V, 10 V, 100 mV 60 V only 10 V only 100 mV DC Voltage Source 10 or 60 V DC Voltage Source NOTE Maximum input voltage to housing and supply ground is ±60 V. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 109: Current-Fed Piezoelectric Transducers (Iepe, Icp®)

    PCB Piezotronics.   Transducer MX840B-R MX1601B-R MX411B-R Current-fed piezoelectric transducer † † †     ●        ●        ●    (IEPE, ICP®) † Use ODU 14-pin to BNC adapter 1-KAB430-0.3. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 110: Current Sources

    5   SomatXR hardware NOTE IEPE transducers with TEDS version 1.0 are supported. 5.4.7   Current sources   Transducer MX840B-R MX1601B-R MX411B-R ● ● ● Current, 20 mA 20 mA DC Current Source a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 111 5   SomatXR hardware NOTE Maximum current is ±30 mA. 20 mA Current-fed DC Current Source NOTE Maximum current is ±30 mA. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 112: Resistance-Based Measurements

    The sensor supply must be connected in series. This, however, terminates the electrical isolation to the module supply. 5.4.8   Resistance-based measurements   Transducer MX840B-R MX1615B-R Resistance ● ● or resistance-based measurements ● Resistance thermometer (RTD), ● PT100 or PT1000 PT100 only Resistance a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 113: Thermocouples

    5   SomatXR hardware Resistance Thermometer (RTD) 5.4.9   Thermocouples   Transducer MX840B-R MX1609KB-R ● ‡ Thermocouple     ●    K-type only ‡ Use thermocouple adapter 1-SCM-R-TCK-2 for K-type, other types need a dedicated adapter. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 114: Digital Inputs

    5   SomatXR hardware 5.4.10   Digital inputs   Transducer CX23-R ● Digital input, Frequency, Pulse counter, Duty cycle Single-pole, double-throw switch a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 115: Digital Outputs

    5   SomatXR hardware Single-pole, single-throw switch Driven signal 5.4.11   Digital outputs   Transducer CX23-R ● Digital output CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 116 5   SomatXR hardware Sensing circuit Load with external power source Load powered by CX23-R a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 117: Digital Timer Inputs

    Frequency / pulse counter ● (timer, TTL) connectors 5-8 ● ● Incremental encoder (timer, TTL) connectors 5-8 ● ● Torque / speed connectors 5-8     ●    SSI protocol   connectors 5-8 Frequency / pulse counter CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 118 5   SomatXR hardware a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 119 5   SomatXR hardware Incremental encoder CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 120 5   SomatXR hardware Torque / Speed (HBM torque transducers) a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 121 5   SomatXR hardware CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 122 5   SomatXR hardware Absolute value encoder (SSI protocol) a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 123: Passive Inductive Encoder

    5   SomatXR hardware 5.4.13   Passive inductive encoder   Transducer MX460B-R ● Passive inductive encoder 5.4.14   Pulse width, pulse duration, period duration (PWM)   Transducer MX460B-R ● PWM - Pulse width, pulse duration, period duration CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 124 5   SomatXR hardware a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 125: Can Devices

    CAN H and CAN L that can be enabled or disabled individually with software. NOTE The optional CAN power provided by the CX23-R is 1 A limited and the same voltage as the input to the CX23-R. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 126 (+), f (+) and Z (+)) or only the negative inputs ( f (-), f (-) and Z (-)).   Check encoder data sheet for output low and high voltages and set thresholds accordingly. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 127 5   SomatXR hardware For applications requiring long cable runs or more than 200K pulses per second, use the differential wiring diagram below. Place the 120-Ohm termination as close to the input as possible to dampen reflections. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 128 5   SomatXR hardware a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 129 200K pulses per second, a 120 termination should be placed as close to the input as possible to dampen reflections.   Check encoder data sheet for output low and high voltages and set thresholds accordingly. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 130 5   SomatXR hardware For applications requiring long cable runs or more than 200K pulses per second, use the differential wiring diagram below. Place the 120-Ohm termination as close to the input as possible to dampen reflections. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 131 5   SomatXR hardware 5.4.17   Crank wheel sensors   Transducer MX460B-R ● Crank wheel sensors CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 132 5   SomatXR hardware a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 133 CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 134   Maximize a Help window to see the Search bar in the header. NOTE Not all of the interface features may be visible depending on privileges set in Profiles > Privileges (Page 145).   a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 135 The Header contains the View buttons, Test Setup titlebar and Quick Link buttons. Click View buttons to show Task pane menu and action panels and associated workspaces. More View button information The View buttons navigate among options in the interface. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 136 Test Setup. On the Test configuration page, select the Test Setup to edit. On the Test control page, select the Test Setup to run. To the right of the selector are test indicators. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 137 Open the search bar to filter lists or grid entries in the Workspace. When a search is active, the icon is orange . Help Open the online help system. Task pane The Task pane changes according to the View selection and Menu item selection. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 138 LEDs information in the status bar. More Status bar information   Firmware version Currently installed firmware version. The most current CPU load reading and load a chart of the last 60 readings. Mouse over the chart to display data points. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 139 In Test configuration, on a channels tab, a red number also shows on the tab, which is the number of channels resulting from the search. While a search is active, the Search icon will change color. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 140 Most columns can be resized by dragging the column border and reordered by dragging the column header. Select views The eye icon to the right in the header allows connection or channel tabs to be a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 141 The system log panel shows only messages considered most useful to a typical user. Though not displayed, all log entries are included when exporting or purging the log file. 6.5   User preferences Each user can change the user preferences to individually customize the web interface. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 142 To override the number of decimals on a per channel basis, set the Decimals   parameter in the Input channels grid.   To override the number of decimals on a Single channel digital meter, set the number of decimal places from the chart menu. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 143   Maintain aspect ratio for video displays: This is selected by default so video displays retain their aspect ratio (such as 4:3) when their size is changed. If this is not selected, the image in video displays may distort when video display size is changed. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 144   Purge logs   Abort setup task for another session   Delete displays view   Header collect change 6.6   Changing user password To change the current user password, select Change password from User options. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 145 Rename and delete SIE data files Download/Extract Download and extract SIE data files Rename, delete, download and extract SIE data files Logs control None No access to System logs View View and export System logs CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 146 The unit comes with one default user with a user name of admin and a password of password . It is recommended to change the admin password for greater security. Adding a new user To add a new user, click New. Enter the desired values for the user parameters and a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 147 To import a database, navigate to the System view and select the Databases menu option. Click Import. An Import database file dialog shows. Select from the three File types: .dbc (Vector CAN database file), .sdbx (HBM sensor database) or .txt (eDAQ CAN database file).
  • Page 148 NOTE Future data processor firmware versions may support more sensors. The HBM sensor database does not need to be re-imported to use newly supported sensors.
  • Page 149 If on a network with a DHCP server, use the HBM Device Manager (on the included USB drive) or contact your IT administrator to find out the IP address. If presented with a certificate warning, choose to proceed.
  • Page 150 IT administrator. Please note that currently only one mail server and one HBM Push ID can be configured at any given time.
  • Page 151 Click Verify ID and an email is sent to the entered Email address, containing a link to activate the HBM Push ID. Follow the link in the email to activate the Push ID. After the HBM Push ID is activated, selected notifications will result in a message sent to the HBM Push service each time the event occurs.
  • Page 152 At least one notification must be selected for the email configuration to be accepted by the unit. Both Email and HBM Push notifications may be sent when a selected event occurs. Deselect a checkbox and a notification for that event will not be sent.
  • Page 153 After making changes, click the Update button to update the system date and time. 6.7.8   Firmware HBM regularly releases new firmware introducing new features and providing device improvements. Download the current firmware release from www.hbm.com/daq- support.
  • Page 154 Rename modules on the Hardware page. Custom module names cannot match the internal module identifier and can contain only letters, numbers, spaces, dashes or underscores. A module name cannot start or end with a space or spaces. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 155 Four user supplied customizable parameters can be used for a setup's source channel configuration. These can be used for any desired purpose and can contain any desired values. Those with a blank value will not show in the single channel editor. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 156 If deselected, the user defined sample rates and digital filters for each channel are used. Sample rate domain Sample rate Filter type Filter frequency Decimal 100 S/s FIR Butterworth 15 Hz Classic 50 S/s IIR Butterworth 10 Hz a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 157 The system reboots with the default auto network settings and logs in the default admin user. NOTE SIE files from previous test runs are not removed as part of the database delete process and these remain in their current location. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 158  1. Power cycle the system if possible. If not possible, reboot the system using the Reboot control option provided for the CX23-R in the Hardware page.  2. If the message is presented after that, contact HBM customer service. 6.8   Test configuration 6.8.1   Test setup...
  • Page 159   Live updates refresh rate: Specify the rate that live data refreshes on the hardware and input channel pages when live updates are running. A value of zero causes no updates to take place. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 160 If information is entered in System configuration > Preferences > Setup > User info 1, 2, 3 or 4 a custom column name is automatically filled at the creation of a new Test setup. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 161 The setup may be saved as an .sxr text file containing the setup configuration in a json format, or as an .xlsx (Excel 2007) file. Exported setups may be imported to another device as is or after modification. Only one test setup can be exported at a time. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 162 5) Exporting a channel defined via TEDS and/or HBM SDB parameter assignment is supported; however the traceability parameters are not currently included in the export (that is, the exported channels will have the proper parameterization, but when reimported, the channel will not show up with an associated TEDS or SDB state icon).
  • Page 163 Unlike the Normal test run mode, the Remote control test mode requires that the SXR test be enabled (initialized) before any test run can be started. See Running a test for details on this. One topic not discussed in that section is suspending the remote control mode. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 164   Flash LED: Click to identify a control unit or MX module by flashing one of its LEDs. For a CX23-R/EXRCPU, the green POWER LED flashes at 2 Hz. Click Stop flashing to resume normal LED behavior. Normal behavior is also resumed when a module is rebooted. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 165 For other channels, use the presented dialog to set the channel parameters. Applying sensor database parameters To apply sensor configurations from imported HBM sensor databases, first expand the Sensors panel. The Sensors panel contains the imported sensor databases and their sensor definitions.
  • Page 166 Select the desired channels from discovered hardware on the network. Select the check box next to a source branch to display its available channels. The channels are displayed in alphabetical order (by connector) and grouped by source. A a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 167 Sync TEDs. Applying sensor database parameters To apply sensor configurations from imported HBM sensor databases, first expand the Sensors panel. The Sensors panel contains the imported sensor databases and their sensor definitions. For very large databases, it may take several seconds to display the database contents for the first time.
  • Page 168 New connector column. Click OK to save the changes. CAUTION Attempting to supply inappropriate values for the device name results in an error and prevents the test from running. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 169 A list may be sorted by a column's contents, including the Collect column in Test configuration. However, while sorting by the Collect column in the All input channels tab, Computed channels will be sorted separately, below other input channels. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 170   Up sampler (Page 65) NOTE The Profile Generator computed channel is for HBM internal use only and is not supported for other users. Adding a computed channel There are two options to add a computed channel.   Click Add to create a new blank computed channel. If adding from the All computed channels tab, select the type of computed channel to add.
  • Page 171 MX878B module. The channel grid interface displays the output channels defined for the current setup. Grid interface (Page 139) for information about showing and hiding columns. The All output channels view displays generic parameters for all defined output channels. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 172 Select the desired channel or channels and click Remove. Output channels cannot be recovered after removal. Changing device hardware Sometimes it is useful to change the hardware to which a group of channels connects. For example, after importing setup configuration file, using the same setup a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 173 DataModes when possible; otherwise, users will need to be patient in regard to the GUI response when assigning or editing any of the DataMode parameters. Available DataModes:   Burst history (Page 67)   Burst message logger (Page 68) CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 174 Memory Consumption Sequential 32-bit float 4 bytes per data point per channel 64-bit float 8 bytes per data point per channel Histogram 64-bit unsigned 8 bytes per bin per channel 6.9   Test and data control a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 175 To create the first chart, select the desired channels from the Channels list. To add a new chart, drag a channel into an empty area of the Display Views panel. Add CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 176   Delete View: Remove the current display view. Alternatively, click the "x" on the view tab.   Import...: Import charts from a selected Test Setup. An empty chart may also be imported. Save changes to commit to the import. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 177 Up scale: This mode will apply the Up scale (positive swing) shunt polarity for all channels where it is supported by the hardware. Down scale: This mode will apply the Down scale (negative swing) shunt polarity for all channels where it is supported by the hardware. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 178 (It is also closed if the web browser is reloaded or refreshed.) Single channel digital meter The single channel digital meter displays the current value of the selected channel. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 179 Warnings and alarms: When the display is created, the values for these fields are based on the values specified in the grid interface / single channel editor for the selected channel. See System channels > Scaling parameters > CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 180 There are several options available from the menu. Click the options menu button:   Freeze: Select to pause the data display. Deselect to resume live data.   Properties: Open the chart properties dialog box. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 181   Axis label format: Set the time format for the X-axis label among options for hours, minutes, seconds or tenths of seconds. Grid   Hide gridlines: Select to hide the chart gridlines. Analog meter The analog meter is an analog gauge display of a single channel. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 182 2/3 the distance to the new value. When the channel value changes, the needle moves smoothly to the new value, i.e., the higher the time constant value, the slower the needle moves in response to value changes. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 183 FFT Parameters NOTE The Hann window function is applied to the sample window when calculating the spectrum.   FFT size: Set the size of the FFT window in number of samples. Frequency Range (X Axis) CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 184 Min/Max chart will show data only from after its opening. Specify appearance in Test Configuration Specify Range min, Range max, Alarms and Warnings min and max in Test Configuration in the Scaling mode section of the Edit channels dialog or the grid interface. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 185 Please refer to video device manufacturer's documentation for video image rotation, which may be limited to 180 or 90 degree increments. 6.9.2   SIE data The SIE data pane displays all SIE data files stored on the connected device. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 186 The XR Download Manager application is provided to circumvent all browsers and provide reliable and robust downloads of multiple SIE files, and as such, is recommended. This application may be downloaded from the HBM website under Software for QuantumX & SomatXR. See Personal computer support applications for Download Manager installation instructions.
  • Page 187 Viewing SIE data (BETA) Several types of data and information can be viewed from SIE files, including strip charts and video. Select the check box next to the desired SIE file and click View. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 188 Zoom in or out on the charts using the mouse scroll wheel. To view more of a chart while zoomed in on a chart, use the mouse pointer to slide the bar on the time scale at the bottom of the viewer window. a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 189 This indication will appear for a short time after the data processor unit's power-up, or after an MX module is attached or power cycled. The module status should change to PTPv2 or Firewire after a short delay. The data CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 190 2.5 minutes of test data will be lost. Typically when this happens, the MX channels phase skew is less than ½ of a sample period. For most testing applications, this is insignificant and because of that, there is a a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 191 Perform this operation only when communication to the data processor unit is not possible and all other attempts to establish communication have been exhausted. Contacting HBM customer service prior to performing this operation is strongly recommended. All SXR setup files will be moved to a recovery location and will not be available when the rescue is completed;...
  • Page 192 To recover the SXR files that were relocated, do the following. When the unit reboots, immediately navigate to the Customer Service option in the System page and select the Download customer support package. Contact HBM customer service for further instructions.
  • Page 193 "somat:datamode_name" DataMode name       Additional metadata provided in v2.0.0 "somat:user_interface_ For all sample rate domains except the HBM Classic domain, 2500 sample_rate" this is the same as "core:sample_rate"       User interface hardware connection name. Does not include "somat:connection"...
  • Page 194 "somat:electrical_range_ Electrical units range max max"       "somat:bridge_type" full Bridge type (full, half, quarter) Bridge wires (3,4,5,6) (applicable to strain gage bridges, "somat:bridge_wires" inductive bridges, and potentiometers) "somat:bridge_resistance" Bridge resistance (Ohms) a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 195 Max signal level (in mV) crossing required for digital input low" channel to transition to logic 0 (false)       "somat:internal_ Boolean option to use internal termination (applicable to CAN true termination" channels, RS485 sensors on MX460 modules, etc.) CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 196   "somat:sensor_id" 430605296858 HBM SDB ID "somat:sensor_calibration_ 12/22/2017 12:00:00 AM HBM SDB sensor calibration date date" "somat:sensor_calibration_ 12/22/2018 12:00:00 AM HBM SDB sensor calibration expiration date period"         Table 1: Scaling parameters Scaling mode Scaling parameters Example...
  • Page 197 SIE tags for channel parameters). As a rule of thumb, this overhead can be estimated as about three times the SXR file size. While this is insignificant for most testing scenarios, it can be CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 198  2. For histogram DataModes, the storage area is computed using the following formulas.  a. For one dimensional histograms: bytes = (bins + 2) * 8  b. For two dimensional histograms: bytes = (bins + 2) * (bins + 2) * 8 a4275-8.0 en    HBM: public CX23-R and EX23-R...
  • Page 199  c. When sequential DataModes such as the Burst History, Peak Valley and Time History with gated triggering are used, the storage rate is not deterministic and in general not feasible to estimate. CX23-R and EX23-R a4275-8.0 en    HBM: public...
  • Page 200 HBM, Inc. Tel. +1 800-578-4260 info@usa.hbm.com measure and predict with confidence...

This manual is also suitable for:

Somatxr ex23-r

Table of Contents