5 Mains Power

56
5.1

Power and frequency requirements

To connect or disconnect the instrument from the AC supply, plug or unplug the
IEC connector from the instrument. The instrument should be positioned to
allow access to the AC connector. The front power switch on the instrument is
not a disconnecting device. When the instrument is connected some power will
be consumed.
See "Connecting power" on page 57 for more details.
The GEN5i uses up to 500 VA and operates from line voltages from 90 V AC
to 264 V AC at 47-63 Hz. The power connection of the GEN5i is via a standard
IEC 320 EN 60320 C14 (male) appliance inlet, 2-pole, 3-wire designed for
250 V at 10 A. The GEN5i AC supply has no replaceable fuses installed. The
power supply uses a build in resettable power shut down protection.
CAUTION
Do not position the GEN5i so that it is difficult to remove the power input
cable.
The GEN5i must be used with a ground connection trough the conductor of the
supply cable. This is to ensure that all electromagnetic Compatibility (EMC)
requirements are met.
GEN5i
I2679-4.0 en