HBM Genesis GEN5i User Manual page 385

Portable, integrated data acquisition system
Table of Contents

Advertisement

Bessel IIR Filter
When Bessel IIR filter is selected, this is always a combination of an analog Bessel anti alias filter and a digital Bessel IIR filter.
Analog anti alias filter bandwidth
Analog anti alias filter characteristic
Bessel IIR filter characteristic
Bessel IIR filter user selection
Bessel IIR filter bandwidth (ωc)
Bessel IIR passband ripple (δp)
Bessel IIR passband (ωp
Bessel IIR filter stopband attenuation (δs)
Bessel IIR filter roll-off
(1)
Measured using Fluke 5700A calibrator, DC normalized
I2679-4.0 en
Figure A.40: Digital Bessel IIR Filter
400 kHz ± 25 kHz (-3 dB)
7-pole Bessel, optimal step response
8-pole Bessel style IIR
Auto tracking to sample rate divided by: 10, 20, 40, 100
User selects divide factor from current sample rate, software then adjusts filter when sample
rate is changed
User selectable from 0.4 Hz to 20 kHz
(1)
0.1 dB
DC to 3.5 kHz @ ωc = 20 kHz
-60 dB
-48 dB/octave
Figure A.41: Bessel IIR 20 kHz Overview and passband flatness
(1)
GEN5i
385