HBM Genesis GEN5i User Manual page 321

Portable, integrated data acquisition system
Table of Contents

Advertisement

Environmental Specifications
Temperature Range
Non-operational (Storage)
Relative humidity
Protection class
Altitude
Shock: IEC 60068-2-27
Vibration: IEC 60068-2-34
Operational Environmental Tests
Cold test IEC 60068-2-1 Test Ad
Dry heat test IEC 60068-2-2 Test Bd
Damp heat test IEC 60068-2-3 Test Ca
Non-Operational (Storage) Environmental Tests
Cold test IEC 60068-2-1 Test Ab
Dry heat test IEC 60068-2-2 Test Bb
Change of temperature test
IEC 60068-2-14 Test Na
Damp heat cyclic test
IEC 60068-2-30 Test Db variant 1
Harmonized standards for CE compliance, according to the following directives
Low voltage directive (LVD): 2006/95/EC
Electromagnetic compatibility directive (EMC): 2004/108/EC
Electrical Safety
Safety requirements for electrical equipment for measurement, control, and laboratory use - General requirements
EN 61010-1 (2010)
EN 61010-2-030 (2010)
Particular requirements for testing and measuring circuits
Electromagnetic Compatibility
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
EN 61326-1 (2006)
EMISSION
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of
EN 55011
measurement
Conducted disturbance: class B; Radiated disturbance: class A
Limits for harmonic current emissions: class D
EN 61000-3-2
EN 61000-3-3
Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems
IMMUNITY
Electrostatic discharge immunity test (ESD);
EN 61000-4-2
contact discharge ± 4 kV/air discharge ± 8 kV: performance criteria B
EN 61000-4-3
Radiated, radio-frequency, electromagnetic field immunity test;
80 to 2700 MHz using 10 V/m, 1000 Hz AM: performance criteria A
Electrical fast transient/burst immunity test
EN 61000-4-4
Mains ± 2 kV using coupling network. Channel ± 2 kV using capacitive clamp: performance criteria B
Surge immunity test
EN 61000-4-5
Mains ± 0.5 kV/± 1 kV Line-Line and ± 0.5 kV/± 1 kV/± 2 kV Line-earth Channel ± 0.5 kV/± 1 kV using coupling
network: performance criteria B
I2679-4.0 en
Operational
0 °C to +40 °C (+32 °F to +104 °F)
-25 °C to +70 °C (-13 °F to +158 °F)
Thermal protection
Automatic thermal shutdown at 85 °C (+185 °F) internal temperature
User warning notifications at 75 °C (+167 °F) (Supported by Perception V6.30 or higher)
0 % to 80 %; non-condensing; operational
IP20
Maximum 2000 m (6562 ft); operational
Operational
Half-sine 10 g/11 ms; 3-axis, 1000 shocks in positive and negative direction
Non-operational
Half-sine 25 g/6 ms; 3-axis, 3 shocks in positive and negative direction
1 g RMS, ½ h; 3-axis, random 5 to 500 Hz
Operational
Non-operational
2 g RMS, 1 h; 3-axis, random 5 to 500 Hz
-5 °C (+23 °F) for 2 hours
+40 °C (+104 °F) for 2 hours
+40 °C (+104 °F), humidity >93 % RH for 4 days
-25 °C (-13 °F) for 72 hours
+70 °C (+158 °F) humidity <50 % RH for 96 hours
-25 °C to +70 °C (-13 °F to +158 °F)
5 cycles, rate 2 to 3 minutes, dwell time 3 hours
+25 °C/+40 °C (+77 °F/+104 °F), humidity >95/90 % RH
6 Cycles, cycle duration 24 hours
GEN5i
321