Shunt verification
To do the actual shunt verification:
1
In Perception go to the Settings sheet.
2
In the task pane select the Sensors in the Shunt Verification section.
3
Select one or more channels.
4
Enter a value for the Warning level as a percentage.
5
Enter a value as Target : the target value is the result of the bridge value,
excitation value and shunt value. Tables exist for commonly used values.
As an example refer to Figure 12.16. In Figure 12.21 on page 286 the value
is used that corresponds to a 350 Ω bridge, 20 kΩ shunt - therefore a
4.337 mV deflection per volt excitation - and 1 volt excitation.
6
Click Verify. This will actually close S6 in Figure 12.9 on page 266 for a
short period of time to measure the deflection.