HBM Genesis GEN5i User Manual page 440

Portable, integrated data acquisition system
Table of Contents

Advertisement

Continuous
Continuous modes supported
Continuous FIFO memory
Maximum recording time
Maximum aggregate streaming rate per
mainframe
Exceeding aggregate streaming rate
440
Standard, Circular recording, Specified time and Stop on trigger
Standard
User starts and stops recording. Automatic recording stop on storage media full.
Circular recording
User specified recording history on storage media. All recorded data stores as quickly as
possible on selected storage media. As soon as selected history time is reached older
recorded data is overwritten. Recording can be stopped by user, or any system trigger.
Specified time
Automatic recording stop after user specified time or on storage media full
Stop on trigger
Automatic recording stop after any system trigger or on storage media full
Used by enabled channels to optimize continuous streaming rate
Until storage media filled, or user selected time or unlimited using circular recording
Determined by mainframe, Ethernet speed, PC storage medium and other PC parameters;
see mainframe datasheet for details
When using a streaming rate selected higher than the aggregate streaming rate of the
system, the continuous memory will act as a FIFO. As soon as this FIFO fills up, the
recording suspends (temporarily no data is recorded). During this period, the internal FIFO
memory is transferred to storage medium. When internal memory is completely empty
again, the recording automatically resumes. User notifications added to recording file for
post recording identification of storage overrun.
GEN5i
I2679-4.0 en

Advertisement

Table of Contents
loading