Do one of the following:
In the spreadsheet style matrix in the Bridge type column select the
l
bridge type Full.
In the simplified graphical diagram click on the bridge icon (D in Figure
l
12.17 on page 273) until you see the full bridge representation.

Bridge completion - half (1/2 or 2/4) bridge

A half bridge type sensor is a sensor that has two bridge resistors on-board,
completion is required.
Figure 12.19: Half bridge layout
In this situation a sensor is used with two (out of four) resistors R4 and R3.
These resistors are placed between A-C and C-D.