Gn811 Basic 1M Input Card - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
12.3.2

GN811 Basic 1M input card

8 analog channels
l
Single ended inputs
l
± 1 V to ± 50 V input range
l
User selectable digital Bessel
l
and FIR filters
1 MS/s sample rate
l
16 bit resolution
l
256 MB memory
l
Single metal BNC for each
l
channel
The GEN DAQ Basic 1 MS/s input
card is a general purpose signal
conditioner for use with voltage
inputs, externally conditioned
signals or probes and current
clamps.
The basic signal conditioner
provides eight channels of single
ended voltage inputs from ± 1 V to
± 50 V full scale with full offset and
auto-zero capability. Every channel
is equipped with an independent full
range input amplifier, 7-pole Bessel
and Butterworth anti-alias filter, 16-
bit Analog-to-Digital converter
operating at 1 MS/s and several
selections of digital filtering.
The on-board transient memory size
is 128 Mega-Samples (256 Mega-
Bytes). The memory is shared
among enabled channels.
Each channel also features two set-points for trigger or alarm purposes.
Extensive acquisition and trigger modes allow many different ways to capture
valuable data even at the highest sample rates. All channels are synchronously
sampled at full speed without multiplexing and almost immeasurable crosstalk.
The model uses standard metal BNC connectors. The inputs are 1 MΩ
impedance and are compatible with probes and current clamps.
For more information about the GN811 Basic 1M input card, please refer to
"B2640-3.0 en (GEN series GN811)" on page 416.
GEN5i
259

Advertisement

Table of Contents
loading

Table of Contents