Trigger Modes; Basic Trigger Mode - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
10.3

Trigger modes

Using the various trigger modes, your GEN5i data acquisition system is
expanded to an extremely versatile transient recorder. The trigger circuits may
be configured to trigger on many types of phenomena. In this section the
different trigger modes and their extensions are discussed in detail.
10.3.1

Basic trigger mode

The basic trigger mode can be compared with the trigger mode available when
using an analog trigger detector, for example as found on a classic scope.
Figure 10.3: Basic trigger mode
A
Trigger
In this mode a single-level trigger detector is active: the primary level. As
mentioned previously, the signal must actually cross the preset level. Both level
and direction of crossing are selectable.
Relevant settings for this mode:
Mode: Basic
l
Primary level: any value within the input range
l
Direction: positive or negative
l
Hysteresis: any relevant value
l
GEN5i
127