Im1 - Expansion Slot - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
B
External Event Out (EV OUT)
C, D External Trigger In/Out (TRG IN/TRG OUT)
The controller/interface provides 4 BNC connectors with the following functions:
A
External Time base In (TB-A IN)
This input can be used to provide another time base for the ADC rather
than the internal one. Typically used in combination with rotating
machinery where the ADC clock is synchronized with the revolutions. In
the Perception software the selection between external and internal time
base is made in the Mainframe section of the Settings.
B
External Event Out (EV OUT)
This output is software selectable between Alarm Out and Recording
Active Out. When alarm is selected, the output is driven by channel
alarm detectors. When recording active is selected, the output is "high"
when a recording is in progress.
C, D External Trigger In/Out (TRG IN / TRG OUT)
This input and output are related to the recorder trigger logic. For details
see "Recorder and system trigger" in "GEN series Data Acquisition"
manual.

IM1 - Expansion slot

11.4.3
Figure 11.29: Free slot on Interface/Controller Module IM1
A
Free space for option
GEN5i
179

Advertisement

Table of Contents
loading

Table of Contents