276
You now have to provide the two additional resistors R1 and R2. To do this
there are two options:
1
Use the standard 100 kΩ resistors inside the acquisition card.
2
Provide two resistors.
In situation (1) you do not need to do anything from a hardware point of view.
In situation (2) you will need to add the two resistors to the plug-in module on
the locations marked HB+ and HB-. You will also need to remove the jumper
J1. Refer to Figure 12.9 on page 266 for electrical/schematic details and to for
mechanical/location details.
To connect such a bridge you need a minimum of three wires. Refer to Figure
12.14 for connection details. When using a half bridge you need to inform the
Perception software about this.
To select half bridge completion in Perception
To select half bridge completion in Perception proceed as follows:
1
In Perception go to the Settings sheet.
2
In the task pane select the Bridge in the Input section.
3
Select one or more channels.
GEN5i
I2679-4.0 en