HBM Genesis GEN5i User Manual page 276

Portable, integrated data acquisition system
Table of Contents

Advertisement

276
You now have to provide the two additional resistors R1 and R2. To do this
there are two options:
1
Use the standard 100 kΩ resistors inside the acquisition card.
2
Provide two resistors.
In situation (1) you do not need to do anything from a hardware point of view.
In situation (2) you will need to add the two resistors to the plug-in module on
the locations marked HB+ and HB-. You will also need to remove the jumper
J1. Refer to Figure 12.9 on page 266 for electrical/schematic details and to for
mechanical/location details.
To connect such a bridge you need a minimum of three wires. Refer to Figure
12.14 for connection details. When using a half bridge you need to inform the
Perception software about this.
To select half bridge completion in Perception
To select half bridge completion in Perception proceed as follows:
1
In Perception go to the Settings sheet.
2
In the task pane select the Bridge in the Input section.
3
Select one or more channels.
GEN5i