Interval Timer - More - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
Figure 10.12: Interval timer - Less
A
Trigger
B
Interval
The time interval is reset on the first new trigger event. This feature allows you
to detect additional pulses in a standard train of pulses for example.

Interval timer - More

This interval timer mode is more complicated. When the second trigger event
is within the set time interval, no trigger is generated and the time interval is
reset on each trigger event. When a new trigger event occurs after the specified
time interval, i.e. the interval is not reset in time, then a trigger is generated at
the end of the specified time interval.
In the reset moments are denoted with a dotted line, the actual trigger moment
with a straight line.
Figure 10.13: Interval timer - More
A
Trigger
B
Interval
GEN5i
137