As explained later in this chapter, a signal must actually cross the preset level.
This is to avoid erroneous triggering on a small amount of noise on the signal.
To make the trigger detector even more stable when noisy signals are used,
the single-level trigger detector has been expanded with a hysteresis. This
basically doubles the logic.
For the advanced trigger modes the single-level trigger detector with
programmable hysteresis has been implemented twice to provide a dual-
level trigger detector with selectable hysteresis on each channel. Levels are
usually referenced as primary trigger level and secondary trigger level.
124