Figure 12.17: Perception Bridge Amplifier simplified block diagram
A
Channel select
B
Start Bridge Wizard
C
Legend
D
Bridge representation (click to toggle bridge completion)
E
Sense on/off (S1a and S1b in Figure 12.9 on page 266)
F
Show connector pinning on/off
G Excitation type
H
Excitation value
I
Excitation on/off (S2a and S2b in Figure 12.9 on page 266)

Bridge completion

The Wheatstone bridge used in most strain gage measurement circuits usually
consists of (a) the gages for actively measuring the strains and (b) precision
resistors for completing the circuit. In the GN410 and GN411 bridge completion
can be for full, half and quarter bridge configurations. Completion resistors can
be internal (incorporated in the GN410 and GN411) or external (when required).

Bridge completion - full (4/4) bridge

A full bridge type sensor is a sensor that has all four bridge resistors on-board,
no completion is required.