HBM Genesis GEN5i User Manual page 358

Portable, integrated data acquisition system
Table of Contents

Advertisement

Analog Input Section
Common Mode (referred to system ground)
Rejection Ratio (CMRR)
Input overload protection
Over voltage impedance change
Maximum nondestructive voltage
Maximum overload without auto range
Automatic auto range
Overload recovery time
Isolation
Channel-to-chassis (earth)
Channel-to-channel
(1)
IEC61010-1 Category voltage ratings are RMS voltages.
(2)
Channel to Channel CAT II and CAT III ratings are not a valid method to specify.
358
Less than or equal to ± 5 V
Ranges
> 80 dB @ 80 Hz (-110 dB typical)
Voltage
7 V RMS
Figure A.22: Typical Common Mode Rejection
The activation of the over voltage protection system will result in a reduced input impedance.
The over voltage protection will not be active as long as the input voltage is less then
200 % of the selected input range or 1250 V whichever is the smallest value.
± 2000 V DC
200 % of selected range
When overload creates over heating of the amplifier, the amplifier will change its range up
in steps of factor 10 until overload disappears. When the actual overload is above 1000 V,
the input signal will be disconnected and the amplifier input will be grounded. When
temperature returns to normal the original selected range will be restored. Automatic auto
range can not be turned off.
Restored to 0.1 % accuracy in less then 5 μs after 200 % overload
1000 V RMS
2000 V RMS
Larger than ± 5 V
> 60 dB @ 80 Hz (-80 dB typical)
1000 V RMS
CAT II
(1)
600 V RMS
(2)
GEN5i
CAT III
(1)
300 V RMS
(2)
I2679-4.0 en