In Slave mode:
the top connector is configured as slave input, all other connectors are
l
unused
all received signals are transferred to a bus for internal distribution
l
internal slave trigger signals are transferred to the outside
l
In Stand-alone mode:
Stand-alone mode is OFF: the Master/Slave card does not communicate
l
with other Master/Slave cards.
Fiber optic cable
The Master/Slave card has optical I/O (IN/OUT) that connects to other Master/
Slave cards.
The fiber optic cable:
allows up to a 500 m cable between the master and each slave
l
distributes the sampling clock, absolute time info, trigger and start/stop
l
signals between mainframes, creating a real high channel synchronized
system out of the nine mainframes
enables a timing accuracy between the mainframes less than 100 ns
l
11.7.1

Master/Slave card operations

The Master/Slave card is easily inserted into the GEN series mainframe and is
automatically recognized by the Perception software. A Master/Slave card can
be used as a master or slave. One card is required in the master mainframe
and one card is required per slave mainframe. Each card has eight LC
connectors for the fiber optic cable.
GEN5i
®