A.16

B3245-1.0 en (GEN series GN6470)

Analog to Digital Conversion
Component
Sample rate
Type
Digital Input Section
Component
Connectors
Type
Pull-up
Voltage range
Voltage levels
Protection
Reset modes
External
Modes
Conditional functionality
Modes
Trigger
Qualifier
Alarm
Functionality
General
496
Unit Description
Maximum
TTL, active low with pull-up resistor to enable activation by relays or
short-circuit to ground
Unit Description
Four connectors with 16 events per connector
TTL
TTL compatible, maximum
Logic threshold voltage level '0'
Logic threshold voltage level '1'
Logical
Manual (software control), On Start of Acquisition, Use an External
Trigger Once, Use an External Trigger Always. The reset functionality
of the counter/ timer is under software control and can be set for each
channel separately.
Trigger, Qualifier, Alarm
Modes
Off, rising edge active, falling edge active
Combination
Each event trigger is OR-ed with all other trigger
sources
Modes
Off, active high/low
Combination
Each event qualifier is AND-ed with all other
qualifier sources
Modes
Off, active high, active low
Number of Channels
Pins per channel
Function
Clock
l
Direction
l
Reset
l
Figure A.103: Logic threshold voltage levels
GEN5i
Value
1 MS/s
Value
4 x 26-pin SubD
KF66-A26P-N
25.5 kΩ @ 5 Volt
30 V
- 30 V to + 0.7 V
30 V Maximum
± 30 V continuous
"1" or "0"
3
3
I2679-4.0 en