HBM Genesis GEN5i User Manual page 322

Portable, integrated data acquisition system
Table of Contents

Advertisement

Harmonized standards for CE compliance, according to the following directives
Low voltage directive (LVD): 2006/95/EC
Electromagnetic compatibility directive (EMC): 2004/108/EC
EN 61000-4-6
Immunity to conducted disturbances, induced by radio-frequency fields
0.15 to 80 MHz, 1000 Hz AM; 10 V RMS @ mains, 3 V RMS @ channel, both using clamp: performance criteria A
Voltage dips, short interruptions and voltage variations immunity tests
EN 61000-4-11
Dips: performance criteria A; Interruptions: performance criteria C
IRIG, IRIG/GPS (options, to be ordered separately; available Q3 2014)
(1)
IRIG
Supported by IRIG and IRG/GPS option
Time Code Translator (Input)
Time Code Output
AM modulated input/output connectors
DCLS connector
Time synchronization accuracy
GEN DAQ series functions
Time required to full synchronization after IRIG signal detected
Recording or pause active
User notifications while recording
Short term tracking stability
Long term tracking "Fly-wheeling"
(1)
GPS
Only supported by IRG/GPS option
GPS connector
GPS antenna
GPS antenna cable
Time synchronization accuracy
GEN DAQ series functions
GPS localization time
Time required to full synchronization after GPS localization completed
Recording or pause active
User notifications while recording
Short term tracking stability
Long term tracking "Fly-wheeling"
(1)
Requires factory installation
322
Time Code formats
IRIG A and IRIG B, IEEE 1344 compliant
AM Modulated or DC level shift (DCLS)
Modulation ratio
3:1 to 6:1
Input amplitude
500 mV to 5 V Peak-to-Peak
Input impedance
>10 kΩ
Time Code format
IRIG B, IEEE 1344 compliant
Modulation ratio
3:1
Output amplitude
4 V Peak-to-Peak (fixed) into 50 Ω
DC level shift
TTL/CMOS
2 SMB sockets; one for input and one for output
Micro DP, 15-pin; some signals internally linked to Interface/Controller Module
<5 µs modulated, <1 µs (DCLS)
Capture start of recording time
Synchronize Master Time Base oscillator frequency
No recording active
1 to 5 minutes
1 to 5 minutes plus 25 s per ms recording time deviation from IRIG time
Time marks on IRIG signal lost/restored and IRIG time synchronized
5.0 E-8
5.0 E-7
Micro DP, 9-pin
1; included
50 m (164 ft); included
<1 µs
Capture start of recording time
Synchronize Master Time Base oscillator frequency
2 to 15 minutes
No recording active
1 to 10 minutes
1 to 10 minutes plus 25 s per ms recording time deviation from IRIG time
Time marks on GPS satellites lost/restored and GPS time synchronized
5.0 E-8
5.0 E-7
GEN5i