Basic Amplifier None Isolated Input Cards; Gn810 Basic 200K Input Card - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

258
12.3

Basic amplifier none isolated input cards

12.3.1

GN810 Basic 200K input card

8 analog channels
l
Single ended inputs
l
± 1 V to ± 50 V input range
l
User selectable digital Bessel
l
and FIR filters
200 kS/s sample rate
l
16 bit resolution
l
128 MB memory
l
Single metal BNC for each
l
channel
The GEN DAQ Basic 200 kS/s input
card is a general purpose signal
conditioner for use with voltage
inputs, externally conditioned
signals or probes and current
clamps.
The basic signal conditioner
provides eight channels of single
ended voltage inputs from ± 1 V to
± 50 V full scale with full offset and
auto-zero capability. Every channel
is equipped with an independent full
range input amplifier, 7-pole Bessel
and Butterworth anti-alias filter, 16-
bit Analog-to-Digital converter and
several selections of digital filtering.
The on-board transient memory size
is 64 Mega-Samples (128 Mega-
Bytes). The memory is shared
among enabled channels.
Each channel also features two set-
points for trigger or alarm purposes. Extensive acquisition and trigger modes
allow many different ways to capture valuable data even at the highest sample
rates. All channels are synchronously sampled at full speed without multiplexing
and almost immeasurable crosstalk. The model uses standard metal BNC
connectors, whose shells are connected to ground. The inputs are 1 MΩ
impedance and are compatible with probes and current clamps.
For more information about the GN810 Basic 200K input card, please refer to
"B2632-3.0 en (GEN series GN810)" on page 403.