HBM Genesis GEN5i User Manual page 382

Portable, integrated data acquisition system
Table of Contents

Advertisement

Analog Input Section
Channels
Connectors
Input type
Input coupling
AC coupling frequency
Impedance
Ranges
Offset
DC Offset error
DC Gain error
Maximum static error (MSE)
RMS Noise (50 Ω terminated)
382
6
Fully isolated 4 mm banana plugs (Plastic), 2 per channel (red and black)
Analog isolated balanced differential
Coupling modes
AC, DC, GND
48 Hz ± 5 Hz (-3 dB)
Figure A.38: Typical AC coupling response
2 * 1 MΩ ± 1 % // 33 pF ± 10% ranges larger than ± 5 V. All other ranges 57 pF ± 10%
± 20 mV, ± 50 mV, ± 0.1 V, ± 0.2 V, ± 0.5 V, ± 1 V, ± 2 V, ± 5 V, ± 10 V, ± 20 V, ± 50 V,
± 100 V, ± 200 V, ± 500 V, ± 1000 V
± 50 % in 1000 steps (0.1 %);
± 1000 V range has fixed 0 % offset
0.02 % of Full Scale ± 400 μV
Wideband
0.02 % of Full Scale ± 10 μV
All IIR filters
±(20 ppm + 10 μV)/°C (±(12 ppm + 6 μV)/°F)
Offset error drift
0.1 % of Full Scale ± 20 μV
Wideband
0.1 % of Full Scale ± 10 μV
All IIR filters
±30 ppm/°C (±17 ppm/°F)
Gain error drift
0.075 % of Full Scale ± 400 μV
Wideband
0.075 % of Full Scale ± 10 μV
All IIR filters
0.035 % of Full Scale ± 50 μV
Wideband
0.035 % of Full Scale ± 50 μV
All IIR filters
GEN5i
I2679-4.0 en

Advertisement

Table of Contents
loading

Table of Contents