Harmonized standards for CE compliance, according to the following directives
Low voltage directive (LVD): 2006/95/EC
Electromagnetic compatibility directive (EMC): 2004/108/EC
EN 61000-4-6
Immunity to conducted disturbances, induced by radio-frequency fields
0.15 to 80 MHz, 1000 Hz AM; 10 V RMS @ mains, 3 V RMS @ channel, both using clamp: performance criteria A
Voltage dips, short interruptions and voltage variations immunity tests
EN 61000-4-11
Dips: performance criteria A; Interruptions: performance criteria C
Ordering Information
Article
Basic200k
(1)
All GEN series systems are intended for exclusive professional and industrial use.
I2679-4.0 en
(1)
Description
8 Channel, 16 bits, 200 kS/s, ± 1 V to ± 50 V input range,
128 MB RAM (8 MS/channel), single ended, with single
metal BNC for each channel
GEN5i
Order No.
1-GN810-2
415