Acquisition

Since many of the features that are described here are controlled from within
the Perception software, it is advised to read this section in combination with
the corresponding sections in the Perception manual.
The GEN5i/Perception combination provides the following acquisition controls:
RUN The run command starts acquisition of data. Now the recorder(s)
acquire(s) data until a stop command is issued. This stop command can
be manually or triggered (when in sweep storage mode).
STOP To stop or abort an acquisition. The current recording will be closed.
PAUSE This mode has two options:
1
When no acquisition is active it will place the recorder in the pause or
stand-by mode. Although the recorder is digitizing, no data is stored in
memory or disk. This is useful for monitoring purposes.
2
When a continuous acquisition is active, it will place the recorder in a
hold mode: although the recorder is digitizing, no data is stored in
memory or disk. At this point when RUN is selected, the current
recording continues, when STOP is selected, the recording is finished.
These acquisition controls are combined with the various storage modes.