HBM Genesis GEN5i User Manual page 24

Portable, integrated data acquisition system
Table of Contents

Advertisement

24
Figure 2.1: Category indication according to IEC 61010-2-030:2010
Example: A measurement device is specified as 600 V CAT II, maximum input
voltage 1000 V DC.
Table 2.1: Insulation test voltages according to IEC 61010
Nominal Voltage
IEC 61010-2-030:2010
(V RMS or V DC)
5 sec. AC test (V RMS)
CAT II
≤ 150
840
> 150 ≤ 300
1.390
> 300 ≤ 600
2.210
> 600 ≤ 1 000
3.310
Using the above table one can deduct that this specification informs the user
the device passed the insulation tests; 5 sec at 2.210 V RMS and impulse
4.000 V. The maximum operating input voltage is 1000 V DC. This device is to
be used to measure CAT II circuitry up to 600 V.
WARNING
Measurement inputs of this instrument should not be used to measure
high-energy signals of measurement categories CAT II, CAT III or
CAT IV (IEC 61010-2-30:2010) (e.g. mains measurements) , unless
specifically stated for the specific input.
Impulse test (V)
CAT III
CAT IV CAT II
1.390
2.210
1.550
2.210
3.310
2.500
3.310
4.260
4.000
4.260
6.600
6.000
GEN5i
- 2 - 030:2010
CAT III
CAT IV
2.500
4.000
4.000
6.000
6.000
8.000
8.000
12.000
I2679-4.0 en