Analog input GN110, GN111, GN112 and GN113 (Transmitter)
Channels
Connector
Input type
Input Coupling
AC coupling frequency
Impedance
Ranges
DC Offset error
DC Gain error
Maximum static error (MSE)
RMS Noise (50 Ω terminated)
Common Mode (referred to ground and protective ground not connected)
Requires a protected LAB environment and EN50191:2000 compliant work procedures
Rejection Ratio (CMRR)
Input bias current
Rise time
I2679-4.0 en
1
1; Metal BNC
Isolated, unbalanced differential inputs (BNC connected to isolated common)
Coupling modes
AC / DC / GND
1.6 Hz (±10 %); - 3 dB
Figure A.6: Typical AC coupling response
1 MΩ (± 2 %) // 38 pF (± 5 %)
± 20 mV, ± 50 mV, ± 100 mV, ± 200 mV, ± 500 mV, ± 1 V, ± 2 V, ± 5 V, ± 10 V, ± 20 V,
± 50V and ± 100 V
± 50 % in 1000 steps (0.1 %)
Offset
± 100 V range has fixed 0 % offset
0.1 % of Full Scale ± 50 μV
Wideband
0.1 % of Full Scale ± 50 μV
Bessel filter
GN110 and GN111: ±(60 ppm + 10 μV)/°C (±(36 ppm + 6 μV)/°F)
Offset error drift
GN112 and GN113: ±(100 ppm + 10 μV)/°C (±(60 ppm + 6 μV)/°F)
0.1 % of Full Scale ± 50 μV
Wideband
0.1 % of Full Scale ± 50 μV
Bessel filter
GN110 and GN111: ±100 ppm/°C (± 60ppm/°F)
Gain error drift
GN112 and GN113: ±(100 ppm + 10 μV)/°C (±(60 ppm + 6 μV)/°F)
0.1 % of Full Scale ± 50 μV
Wideband
0.1 % of Full Scale ± 50 μV
Bessel filter
0.05 % of Full Scale ± 100 μV
Wideband
0.05 % of Full Scale ± 100 μV
Bessel filter
> 72 dB @ 80 Hz (GN110 and GN111: > 100 dB typical)
Voltage
1.8 KV RMS (GN112 and GN113)
>1.8 kV RMS (GN110 and GN111); Limits set by fiber cable and transmitter air gap isolation
< 2 nA
14 ns