Excitation - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
4
Do one of the following:
In the spreadsheet style matrix in the Bridge type column select the
l
bridge type Quarter.
In the simplified graphical diagram click on the bridge icon (D in Figure
l
12.17 on page 273) until you see the quarter bridge representation.
5
When you select quarter bridge completion, switch S3 in Figure 12.9 on
page 266 will be closed.

Excitation

The following options are provided for bridge excitation:
You can switch excitation on and off.
l
You can select between voltage and current excitation.
l
You can use sense lines to make sure that the correct voltage is applied to
l
the bridge even with longer lead wiring.
You make all these selections in the Perception application. However, when
using sense lines you will need to do additional wiring:
Add a connection from pin 3 (+sense) to the bridge connection marked A
l
in Figure 12.9 on page 266.
Add a connection from pin 4 (-sense) to the bridge connection marked D in
l
Figure 12.9 on page 266.
GEN5i