Artificial Star Adapter (option, to be ordered separately)
The artificial star adapter creates an artificial star point for measuring 3 phase signals
Maximum input voltage
Inputs
Outputs
Artifical star N
Safety
Application use
Weight
Material housing
Set-up
Temperature range
Operational temperature
Non-operational (storage)
378
1000 V DC (707 V RMS) between every of the phases
3; 4 mm safety banana plugs
6; 4 mm safety banana pins; plugs straight into GN610/GN611 cards
Reference plug only. Not to be used as input
Conform IEC61010-1 600 V RMS CAT II
The 3 phase signals L1, L2 and L3 can be connected with inputs L1, L2, L3 of the artificial
star adapter. The connection N* is the voltage present on the artificial "star point".
Figure A.34: Typical use of artificial star adapter
Figure A.35: Electrical schematic
170 g
Poly Urethane, vacuum resin casting
2 boxes can be plugged onto a single GN610/GN611 card
2 GN610/GN611 cards with Artificial star adapters will fit next to each other
0 °C to +40 °C (+32 to +104 °F)
-25 °C to +70 °C (-13 °F to +158 °F)
GEN5i
I2679-4.0 en