B
Verify command
C
Warning settings
D
Channel select
E
Schematic block diagram of balance circuit
To balance a bridge
To balance a bridge in Perception do the following:
1
In Perception go to the Settings sheet.
2
In the task pane select Bridge Balance in the Sensors section.
3
Use the Balance Enable column to enable/disable the balancing of
channels.
4
Enter a value for the Warning level as a percentage.
5
Select one or more channels.
6
Click the Balance command and wait for the results.