CAUTION
If the end point potentiometer is turned, it changes the factory–set
calibration for the voltage output.
Fig. 7.3: Setting up voltage output zero point
7.4 Functional testing
7.4.1 Power transmission
If you suspect that the transmission system is not working correctly, you can
remove the stator cover and check the functioning of the equipment. If the
LED is shining, the rotor and stator are properly aligned and there is no
interference with the transmission of measuring signals. When the calibration
signal is triggered, the LED shines more brightly.
Fig. 7.4: Checking that power transmission is functioning correctly
B 23.T10F.90 en
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
35
End point
Zero point
Test-LED