Back Manifold Plate

Thermo Scientific
To replace the back manifold plate
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or column. However, you must
vent the instrument.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the power cables from the back of the ISQ instrument.
Figure 260. Disconnecting the Power Cables
Screws
Power Cable
Connectors
4. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
5. Grasp the power supply system and lift it out of the instrument. It fits snug, so you may
have to shake it as you lift it out
3
Advanced Maintenance
Replacing Components of the Manifold
"Shutting Down the ISQ 7000 System"
ISQ 7000 Hardware Manual
249