Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
282
ISQ 7000 Hardware Manual
6. Attach the new chassis feet. The dimensions are in
Figure 296. Dimensions of the Feet on the Bottom of the ISQ 7000 Instrument
4.5 in (12 cm) from
front foot to front of
instrument with door
closed
Tip Thermo Fisher Scientific recommends that all four of the instrument's feet be
positioned on a bench. Overhang of the front, the back, or both ends of the instrument
off the bench front or back is acceptable. If the depth of your bench is less than the
distance between the factory-installed feet, please contact the factory
(usaus.gcms.licensing) for your other options of positioning the instrument on such a
narrow bench.
7. Put the ISQ 7000 instrument back into place.
8. Follow the instructions in the section,
components you removed.
9. Follow the instructions in
10. Restart your system by following the instructions in
Figure
15.5 in (40 cm)
24 in. (61 cm) Bottom
of Instrument
"Distribution Board"
"Replacing the Analyzer Tray"
296.
3 in (8 cm) from back
foot to back of
instrument
on
page 323
to replace the
on
page
162.
"Restarting the System"