Distribution Board

Thermo Scientific
The distribution board lies on the bottom of the ISQ 7000 instrument's chassis.
To replace the distribution board
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. Use a T20 Torxhead screwdriver to remove the two screws holding the stainless steel
support bracket to the middle of the instrument.
Figure 351. Removing the Support Bracket
3
Replacing the Boards in the ISQ 7000 Mass Spectrometer
"Shutting Down the ISQ 7000 System"
Support Bracket
ISQ 7000 Hardware Manual
Advanced Maintenance
323