Transfer Line

Thermo Scientific
When replacing the transfer line, it is important to properly connect the nut to the transfer
line. Otherwise, you will get inconsistent or unreliable data because air or other contaminants
will get inside and interfere with the performance of the system.
IMPORTANT Be sure to install the transfer line correctly or the ion source will not align
correctly.
To replace the transfer line and ferrule
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
149.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Figure 414. Removing the Right Side Panel Screws
4. Slide the right side panel to the back of the instrument and lift it off.
Replacing the Calibration Gas Components
"Shutting Down the ISQ 7000 System"
Right Side Panel Screws
ISQ 7000 Hardware Manual
3
Advanced Maintenance
365