Advanced Maintenance
Replacing Components of the Manifold
Source Retainer Magnets, Magnet Yoke, Source Insertion Guide, Source Magnets
and Front Manifold O-ring
252
ISQ 7000 Hardware Manual
To replace the source retainer magnets, magnet yoke, source insertion guide, source
magnets, or front manifold o-ring on the manifold door
Note For information about replacing the tray alignment pin or the o-ring on the back
manifold plate, see
Back Manifold
Note For information about ordering these components, see the ISQ and TSQ GC-MS
Spare Parts Guide.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 263. Opening the Manifold Door
Plate.
"Shutting Down the ISQ 7000 System"
Screws
Thermo Scientific