Front Door Latch

Thermo Scientific
To replace the latch on the front door
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Slide the left side panel toward the back of the instrument and lift it off.
4. Use a T10 Torxhead screwdriver to remove the two screws holding the latch onto the
front door.
Figure 302. Removing the Latch from the Front Door
Screws
3
Advanced Maintenance
Replacing Components of the Front Door
"Shutting Down the ISQ 7000 System"
Latch
ISQ 7000 Hardware Manual
287