To replace the insulating spacers
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 182. Opening the Manifold Door
3. Twist off the source-repeller thumbscrews one at a time to prevent the repeller plate from
falling out.
Figure 183. Removing the Source-Repeller Thumbscrews
"Shutting Down the ISQ 7000 System"
Screws
Source-Repeller
Thumbscrews
Thermo Scientific