Thermo Scientific ISQ 7000 Hardware Manual page 94

Mass spectrometers
Table of Contents

Advertisement

1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
66
ISQ 7000 Hardware Manual
Figure 55. Removing the Manifold Door Screws
8. Open the manifold door.
9. Let the system cool down for 30 minutes.
CAUTION - BURN HAZARD The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE : soyez prudent, car la cartouche de la source d'ions
peut être chaude.
10. Disconnect the filament wires from the source interface board.
Figure 56. Disconnect the Filament Wires
11. Grasp the Vespel cap and gently remove the Advanced EI ion source from the ion source
block. See
Figure
57.
Filament Wires
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents