Thermo Scientific ISQ 7000 Hardware Manual page 269

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
9. Replace the manifold door hinge and reattach the screws.
10. Replace the shield and door, making certain that the four screws holding the door to the
manifold are aligned before tightening the three screws holding the new hinge to the
manifold.
11. From the front of the instrument, reattach the vacuum interlock shield.
Figure 247. Reattaching the Vacuum Interlock Shield
12. Align the connector on the bottom of the lens driver board with the connectors on the
distribution board beneath it. It will drop into the connector when its in the right
position.
Figure 248. Reinstalling the Lens Driver Board
Connector
3
Advanced Maintenance
Replacing Components of the Manifold
Vacuum
Interlock
Shield
Springs
ISQ 7000 Hardware Manual
241