15. Insert a new turbo pump into the chassis.
Note There is a centering ring with an o-ring seal that must be correctly placed between
the pump and the vacuum manifold.
Figure 459. Installing the Standard Capacity Turbo Pump
16. Lift the turbo pump up and onto the support clamps.
17. Reattach the remaining clamps.
18. Tighten all the nuts holding the turbo pump to the chassis.
19. Attach the adapter to the turbo pump and clamp it in place.
20. Connect the turbo pump cable to the 15-pin connector on the distribution board.
Figure 460. Reconnecting the Turbo Pump Cable to the Distribution Board
4
Upgrade Equipment
Turbomolecular Pumps
Turbo Pump
Cable
Connector
ISQ 7000 Hardware Manual
399