Thermo Scientific ISQ 7000 Hardware Manual page 225

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
4. Use forceps to remove and replace the thumbscrew-repeller spacers.
Figure 181. Replacing the Thumbscrew-Repeller Spacers
Note When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crush the spacers.
5. Reattach the source-repeller thumbscrews to the spacers.
6. Reattach the front manifold door and four screws.
7. Restart your system by following the instructions in
Replacing Components of the Ion Source
Thumbscrew-Repeller
Spacers
"Restarting the System"
ISQ 7000 Hardware Manual
3
Advanced Maintenance
on
page
165.
197