Top Cover Panel - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer

Top Cover Panel

276
ISQ 7000 Hardware Manual
To replace the top cover panel of the ISQ 7000 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
Figure 290. Replacing the Top Cover Panel of the ISQ 7000 Instrument
Top Cover Panel Screw
3. Replace the top cover panel and tighten the screw holding it in place.
Restart your system by following the instructions in
"Shutting Down the ISQ 7000 System"
Top Cover Panel
"Restarting the System"
on
page
165.