7. Remove and replace the electrometer board.
8. Reattach the screws that hold the electrometer board in place.
9. Reattach the electrometer shield and screws.
Tip To prevent the screws from falling down into the instrument, place them on the shield
before you reattach it to the chassis.
Figure 386. Reattaching the Electrometer Shield
10. Reconnect the electrometer board cable.
11. Reattach the top cover panel and tighten the screw holding it in place.
12. Restart your system by following the instructions in
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Screws
"Restarting the System"
ISQ 7000 Hardware Manual
3
Advanced Maintenance
on
page
165.