29. Reattach the three screws holding the valve body to the front manifold plate.
Figure 236. Reattaching the Valve Body to the Front Manifold Plate
Screws
30. Reattach the two screws connecting the pillow block to the manifold door.
Figure 237. Reattaching the Pillow Block
Pillow
Block
Screws
CAUTION INSTRUMENT DAMAGE: If the vacuum interlock handle is not fully down
and the ball valve is not fully closed, then the ball valve may not be able to completely
open it during use and it could damage the ball valve and probes.
3
Advanced Maintenance
Replacing Components of the Vacuum Interlock
Front
Manifold
Plate
ISQ 7000 Hardware Manual
233