Figure 208. Replaceable Components of the Advanced EI Ion Source Cartridge
Filament
Vespel Cap
AEI Filament
Spacer
Thermo Scientific
AEI Filament
Insulator
AEI Filament
Shield
To replace components of the ExtractaBrite ion source cartridge
1. Follow the procedure in
entire ion source cartridge.
2. Replace the components as necessary.
3. Reassemble and reinsert the ion source cartridge by following the steps in
Ion Source Components"
To replace components of the Advanced EI ion source cartridge
1. Follow the procedure in
System"
on
page 64
to remove the ion source cartridge.
2. Replace the components as necessary
3. Reassemble and reinsert the ion source cartridge by following the steps in
the Advanced EI Ion Source Cartridge"
Lens 3/RF
AEI Ion Volume
Lens
Insulator
Lens 2
AEI Ion Volume
"Removing the Ion Source Cartridge"
on
page
46.
"Removing the Advanced Ion Source Cartridge on a ISQ 7000
on
page
3
Advanced Maintenance
Replacing Components of the Ion Source
AEI Ion Cartridge
Sleeve
on
page 48
82.
ISQ 7000 Hardware Manual
to remove the
"Replacing the
"Reassembling
217