1
Performing Routine Maintenance
Replacing a Dual Filament
94
ISQ 7000 Hardware Manual
CAUTION INSTRUMENT DAMAGE: Use clean, lint-free gloves when touching
components inside the vacuum manifold.
ATTENTION DOMMAGES À L'INSTRUMENT : utilisez des gants propres non pelucheux
lorsque vous touchez des composants à l'intérieur du collecteur à vide.
14. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 92. Opening the Manifold Door
15. Open the front door to move the magnet yoke away from the filament.
Screws
Thermo Scientific