Advanced EI Ion Source

Advanced EI Ion Source
374
ISQ 7000 Hardware Manual
If you did not order a ISQ 7000 system with an Advanced EI ion source, you can order a
system later. You must change the source magnets on the manifold door to use an Advanced
EI source. To install an Advanced EI ion source.
1. Follow the steps in
"Removing the Ion Source Cartridge"
Extractabrite ion source cartridge.
2. Follow the steps in
"Powering Off the ISQ 7000 Mass Spectrometer"
down the ISQ 7000 system.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
4. Open the manifold door.
Figure 423. Opening the Manifold Door
on
page 48
Screws
to remove the
on
page 13
to shut